local · auto-updatingHow long will a CPAP battery really last? Enter the machine's watts, the pack's watt-hours or amp-hours and the path between them. Get the nights of sleep with depth of discharge, converter losses, cold and the FAA carry-on limit counted — or the battery size you need for the trip.
the formula
Nights = (watt-hours × depth of discharge × cold factor) ÷ (machine watts ÷ path efficiency + path idle watts) ÷ hours slept. At the defaults — a 35 W heated machine, a 1,000 Wh lithium bank at 85% usable, through a 12 V to 24 V converter at 87% — that is 850 Wh usable against 41.2 W of draw: 20 h 37 m, about 2.6 nights. On a direct 12 V input the same bank gives 24 h 17 m.
machine + battery + path
Metered on a test lung, not the label figure: ResMed's AirSense 10 measures 7 W bare, 22 W humidified, 34 W with a heated tube.
Average over the night, not the peak.
Everything between the battery and the machine.
Victron quotes 85–89% for a DC-DC box at light load.
What the converter burns doing nothing — an inverter's 6 W matters at CPAP watts.
Ah × volts = Wh. A 100 Ah 12 V battery is 1,200 Wh.
The number printed on the pack.
Usually printed as C20 for lead-acid.
12 V, 24 V or a 25.6 V LiFePO4 box.
Sets the depth of discharge.
A garage, an RV in winter or a tent at altitude.
The night you actually wear the mask.
Judged against this pack here, and used as the target in the size-to-buy mode.
Adds the 150%-of-flight-time check some airlines ask of medical devices.
US average residential, EIA Table 5.6.A — for comparing heat settings.
Real setups
live caveats
These notes update with your inputs, and they are the reasons a CPAP battery claim on a product page rarely survives contact with a real night.
how the math works
Three numbers decide the answer: what the machine really pulls, what the battery really gives, and what the path between them burns. Here is each one with its source.
A CPAP's demand is dominated by its heaters, not its blower. ResMed publishes bench figures measured on a test lung, and they are a fraction of the nameplate: the AirSense 10 AutoSet draws 7 W as a bare machine, 22 W with the humidifier running and 34 W when a ClimateLineAir heated tube is added. The AirCurve 10 VAuto measures 41 W and a CS PaceWave 44 W. The same AirSense 10 carries a label reading 53 W typical and 104 W maximum, because a label has to cover the worst-case heater setting and the highest rated altitude.
| Configuration | Metered watts | Watt-hours per 8 h night | What it means on a 1,000 Wh bank |
|---|---|---|---|
| Travel unit, no heat | 5–10 W | 40–80 Wh | 10–21 nights |
| Pressure only, no humidifier | 7 W | 56 Wh | about 15 nights |
| Humidifier on | 22 W | 176 Wh | about 4.8 nights |
| Humidifier + heated tube | 34 W | 272 Wh | about 3.1 nights |
| Bi-level, maximum heat | 41–55 W | 328–440 Wh | 1.9–2.6 nights |
Last column is 1,000 Wh × 0.85 usable ÷ those watts ÷ 8 h, on a direct 12 V input. Add a converter or an inverter and each figure drops — see the next table.
Three things sit between nameplate watt-hours and energy at the mask. Depth of discharge is the share the pack is designed to give up — 90% for LiFePO4, 85% for the li-ion cells in power banks and stations, 50% for lead-acid and AGM if you want the cycles. Cold takes capacity directly: a pack at 0 C gives up roughly a fifth, and one at −20 C about half. Then the battery management system cuts off before empty, so the last few percent is never available at any voltage the machine can use.
Amp-hour batteries are the common source of error, because an Ah figure means nothing without its voltage. A "100 Ah" battery is 1,200 Wh at 12 V, 1,280 Wh at a 25.6 V LiFePO4 nominal, and 2,400 Wh at 24 V. The calculator multiplies them for you; a vendor quoting nights from amp-hours alone is guessing.
Nothing between the battery and the machine is free. A DC-DC step-up converter for a 24 V ResMed input runs at 85–89% at light load — Victron's published figure for the Orion-Tr range — and draws a little of its own. An AC inverter is worse in both ways: 85% at best on a small load, and a no-load idle draw that a 10 W travel machine cannot outrun.
| Path | Efficiency | Idle draw | Battery watts at 35 W | Nights from 1,000 Wh |
|---|---|---|---|---|
| DC direct into the machine's 12 V input | 100% | 0 W | 35.0 W | 3.04 |
| DC-DC step-up 12 V to 24 V | 87% | 1 W | 41.2 W | 2.58 |
| AC inverter, pure sine | 85% | 6 W | 47.2 W | 2.25 |
| AC inverter at 10 W (travel unit) | 85% | 6 W | 17.8 W | 5.97 |
| DC direct at 10 W (travel unit) | 100% | 0 W | 10.0 W | 10.63 |
Nights = 850 Wh usable ÷ battery watts ÷ 8 h. The last two rows are the same travel machine and the same battery: the inverter's 6 W idle costs you 4.7 nights of the 10.6 available.
The second mode runs the formula backwards. Decide the nights, and the calculator returns the nameplate watt-hours that must be on the box:
That single comparison is the most useful number on the page. Buying into the inverter path costs you 34% more capacity for the same three nights — which is why a 12 V DC cable is the cheapest upgrade in CPAP travel, and why the "100 W CPAP needs a 1,000 Wh battery" advice floating around is usually double what the machine actually wants.
The FAA divides lithium batteries by watt-hours, and the CPAP-legal sizes are small. Under 100 Wh, spare batteries and power banks travel in carry-on with no approval. Between 100 and 160 Wh, they need the airline's approval and are limited to two spares per passenger. Above 160 Wh they are forbidden on passenger aircraft — no exception for a medical device, which is why a 500 Wh power station cannot fly. Several carriers ask CPAP and oxygen users to carry enough runtime for 150% of the flight time, covering delays, so a five-hour flight wants 7.5 hours of run time: at 10 W on a DC input that is 75 Wh usable, about 88 Wh nameplate, which a 96 Wh airline pack clears.
| Pack size | Status on a passenger flight | Typical product at that size |
|---|---|---|
| Up to 100 Wh | Carry-on, no approval needed | Large laptop power bank, 96 Wh CPAP pack |
| 101–160 Wh | Carry-on with airline approval, two spares max | Purpose-built CPAP battery, 150 Wh station |
| Over 160 Wh | Not permitted as a battery or power bank | Any real power station — check it in as device baggage only if installed |
About 20 hours 37 minutes on a 1,000 Wh lithium pack for a machine drawing 35 W — roughly 2.6 eight-hour nights — once you count the 85% usable depth of discharge and the 87% a 12 V to 24 V converter passes. Run the same 35 W on a direct 12 V input and the same pack lasts 24 hours 17 minutes, about 3.0 nights.
ResMed measures its AirSense 10 AutoSet at 7 W on a test lung with nothing heated, 22 W with the humidifier on and 34 W with the humidifier plus a ClimateLineAir heated tube; the AirCurve 10 VAuto measures 41 W. The label figures are much higher — the AirSense 10 is marked 53 W typical and 104 W peak — because they assume maximum heat, so metered watts are the right number for a runtime estimate.
Three eight-hour nights of a 35 W machine delivers 840 Wh on a direct 12 V input, which is 988 Wh of nameplate lithium at 85% usable. Through a DC-DC converter the same three nights cost 990 Wh delivered and 1,164 Wh nameplate; through an AC inverter it is 1,132 Wh delivered and 1,332 Wh nameplate. Turning the heat down moves that number far more than buying a slightly bigger box does.
Yes, in carry-on baggage. The FAA lets spare lithium batteries and power banks up to 100 Wh travel without approval; 101 to 160 Wh are allowed only with the airline's approval and are limited to two spares per passenger; anything above 160 Wh is forbidden on a passenger aircraft. Several airlines also ask CPAP users to carry enough battery for 150% of the flight time, so a five-hour flight wants about 7.5 hours of run time.
Not for a small load. A power station's AC inverter burns about 6 W just sitting switched on and loses 12–15% converting, so at a 10 W travel-machine setting the idle draw alone is about 34% of everything the battery supplies. A 12 V battery feeding the machine's own DC input, or a DC-DC step-up converter, keeps both the watts and the nights. Use a station when you also want to run a lamp and a phone.
It is the biggest single consumer on the machine. ResMed's metered figures go from 7 W with nothing heated to 22 W with the humidifier to 34 W when a heated tube joins them, so the tube heater is roughly the same again on top of the humidifier. Across an eight-hour night that is 56 Wh against 272 Wh — enough to take a 96 Wh airline power bank from about 1.5 nights at 7 W to about 0.3 nights at 34 W.
cpap battery guides
What a CPAP really pulls in watts, a nights-per-watt-hour chart for every pack size, and the setup method for camping, an RV, an outage and a flight.
Metered bench figures from 7 W to 41 W, why the label says 53 W and the peak says 104 W, and how the humidifier decides your battery bill.
96, 150, 300, 500, 1,000 and 2,400 Wh against three machine settings and three power paths — plus the lead-acid and cold columns nobody prints.
The six-step setup for camping, an RV, a power outage and a flight — DC cable versus converter versus inverter, and the FAA carry-on paperwork.